Main Specifications of T-10A
|
|
Model
|
Illuminance Meter
T-10A
(Standard receptor head)
|
Illuminance Meter
T-10MA
(Mini receptor head
|
Illuminance Meter
T-10WsA
(Waterproof mini receptor head)
|
Illuminance Meter
T-10WLA
(Waterproof mini receptor head)
|
Type
|
Multi-function digital illuminance meter with detachable receptor head
(Multi-point measurements of 2 to 30 points is possible)
|
Illuminance meter class
|
Conforms to requirements for Class AA of JIS
C 1609-1:2006 *Illuminance meters Part 1:
General measuring instruments*
Conforms to DIN 5032 Part 7 Class B
|
Conforms to requirements for special
Illuminance meters of JIS C 1609-1:2006
|
Receptor
|
Silicon photocell
|
Relative spectral responsivity
|
Within 6% (f1’) of the CIE spectral luminous efficiency V (λ)
|
Cosine correction characteristics (f2)
|
Within 3%
|
Within 10%
|
Measuring range
|
Auto range (5 manual ranges at the time of analog output)
|
Measuring function
|
Illuminance (lx). Illuminance difference (lx). Illuminance ratio (%). Integrated illuminance (lx· h).
Integration time (h). average illuminance (lx).
|
Measuring
range
|
Illuminance
|
0.01 to 299,900 lx; 0.001 to 29,990 fcd
|
1.00 to 299,900 lx; 0.1 to 29.990 fcd
|
Integrated illuminance
|
0.01 to 999,900 x 103 lx· h 0.001 to 99,990 x 103 fcd· h/ 0.001 to 9999 h
|
User calibration function
|
CCF (Color Correction Factor) setting function: Measurement value x 0.500 to 2.000
|
Linearity
|
+/-2% +/-1 digit of displayed value
|
Temperature/
humidity drift
|
Within +/-3%
|
Measurement speed
|
2 times/sec. (continuous measurement with 1 receptor head)
|
Computer interface
|
USB
|
Printer output
|
RS-232C
|
Analog output
|
1 mV/digit, 3V at maximum reading; Output impedance: 10 KΩ; 90% response time: 28 ms
|
Display
|
3 or 4 Significant-digit LCD with backlight illumination (Automatic illumination)
|
Power
|
2 AA-size batteries/ AC adapter AC-A308 (optional; for 1 to 10 receptors) or
AC adapter AC-A311 (optional; for 1 to 30 receptors)
|
Battery performance
|
72 hours or longer (when alkaline batteries are used) in continuous measurement
|
Operation temperature/
Humidity range
|
-10 to 40๐C, relative humidity 85% or less
(at 35๐C ) with no condensation
|
5 to 40๐C, relative humidity 85% or less
(at 35๐C ) with no condensation
|
Storage temperature/
Humidity range
|
-20 to 55๐C, relative humidity 85% or less
(at 35๐C ) with no condensation
|
0 to 55๐C, relative humidity 85% or less
(at 35๐C ) with no condensation
|
Size (W x D x H)
|
69 x 174 x 35 mm
|
Main body: 69 x 161.5 x 30 mm
Receptor: Ø16.5 x 13.8 mm
|
Cord length
|
-
|
1 m
|
5 m
|
10 m
|
Weight
(without battery)
|
200 g (7.0 oz.)
|
205 g
|
260 g (Receptor head only: 120 g)
|
340 g (Receptor head only: 200 g)
|